Liquid nitrogen safety

On this page you will learn

What liquid nitrogen is

Liquid nitrogen boils at about minus 196 degrees Celsius. It is colourless, odourless and boils constantly at room temperature, giving off nitrogen gas. Every gamete and embryo in the cryo-bank depends on it. It is safe when you respect its three hazards; it is dangerous when you become casual about it.

Hazard 1: extreme cold

Liquid nitrogen or anything cooled by it, such as a metal canister, can cause cold burns within seconds. Splashes into the eye can cause serious injury. Do not touch cooled metal with bare skin. A drop on skin often skates off on a layer of vapour. But liquid trapped in a glove, a cuff or a shoe stays in contact and causes deep burns. That is why gloves must be loose and easy to throw off.

Hazard 2: oxygen depletion

One litre of liquid nitrogen turns into roughly seven hundred litres of gas. In a small or poorly ventilated room, that gas displaces oxygen. You cannot smell or see it, and low oxygen can make a person confused and collapse without warning.

Never work with liquid nitrogen in a closed small room, a cold room or a lift. The cryo-bank room must have good ventilation and a fixed oxygen monitor with an alarm that sounds well before oxygen falls to a dangerous level. Know where the monitor is and what its alarm sounds like.

Hazard 3: pressure

Because liquid nitrogen boils constantly, it must never be sealed in a container that is not designed for it. A closed bottle or a screw-topped container can burst. Use only proper dewars and cryotanks with loose-fitting lids or pressure-relief valves. A closed carrier straw that has let nitrogen in can also crack on warming, so wear eye protection when warming too.

Personal protective equipment

At the bench

Filling and moving

  1. Tell a colleague before you start a fill. Do not handle large volumes alone or after hours.
  2. Put on all protective equipment before you touch the supply vessel.
  3. Fill slowly. Warm containers boil violently at first; let them cool before filling further.
  4. Move dewars on a trolley. A dewar never travels in a lift with a person; send it alone and stop others from using the lift.

Daily tank checks

In the Genesis Day 9 practical you dress in full protective equipment and fill a small dewar under supervision. You also carry out a level check and take part in a labelling, storage and retrieval drill (page 49).

Emergencies

Common errors

Key points

Check yourself

  1. What are the three hazards of liquid nitrogen?

    Extreme cold, oxygen depletion and pressure.

  2. Why must cryo-gloves be loose-fitting?

    So that they can be thrown off quickly. Liquid trapped in a glove stays in contact with the skin and causes deep burns.

  3. Why must a dewar never travel in a lift with a person?

    The nitrogen gas can displace oxygen in a small closed space. You cannot smell or see it, and low oxygen can make a person confused and collapse without warning.

  4. What should you do if the oxygen alarm sounds while a colleague is inside the cryo room?

    Stay out and keep others out. Do not go in to rescue your colleague; call for help and ventilate from outside, because rescuers can be overcome too.

  5. From page 03: List four critical steps that need a witness.

    Any four of: receiving a sample; moving material from one container to another; insemination or ICSI; transfer; placing anything into or removing anything from cryo-storage (page 3).

  6. From page 47: What is the main difference between an open and a closed carrier?

    On an open carrier the drop is plunged directly into liquid nitrogen. In a closed carrier the sample never touches liquid nitrogen (page 47).

Contents and reading plan

Questions on this page

Why is liquid nitrogen dangerous in a small room?

It turns into a large volume of nitrogen gas that pushes out oxygen. Low oxygen cannot be smelled or seen and can cause sudden collapse, so cryo rooms need ventilation and an oxygen alarm.

What should I wear when handling liquid nitrogen?

A face shield or goggles, loose insulated cryo-gloves, a long-sleeved coat or apron over the glove cuffs, closed shoes and long trousers worn over the shoes.

What does frost on the outside of a cryotank mean?

It can mean the vacuum insulation has failed. Report it at once, check the level more often, and be ready to move the samples to a spare tank.

The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Training Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.