Cryoprotectants and carrier devices
On this page you will learn
- The two families of cryoprotectant and how each protects the cell
- Why vitrification uses an equilibration step and then a vitrification step
- How warming solutions reverse the process safely
- The types of carrier device, and how to choose and handle them
Permeating cryoprotectants
Permeating cryoprotectants are small molecules that cross the cell membrane. They replace part of the water inside the cell, lower the temperature at which ice can form and make the solution thicker so that it can form a glass.
The ones you will meet are ethylene glycol, dimethyl sulfoxide (DMSO), propylene glycol (1,2-propanediol) and, for sperm, glycerol. All are toxic in high concentration and at warm temperatures, so vitrification solutions usually combine two agents at lower individual concentrations, and exposure is short and timed.
Non-permeating cryoprotectants
Non-permeating cryoprotectants, mainly sugars such as sucrose and trehalose, stay outside the cell. They draw water out by osmosis, so the cell shrinks and less water is left to freeze. On warming they do a second job: they slow the rush of water back into the cell while the permeating agent leaves, preventing the cell from swelling until it bursts.
A protein supplement or a synthetic polymer is usually added to the base medium to protect the membranes and stop cells sticking to surfaces.
Equilibration and vitrification solutions
A cell cannot be dropped straight into a full-strength vitrification solution: it would shrink too fast and the cryoprotectant would not have time to enter. So vitrification is done in two stages:
- Equilibration solution: a lower concentration of permeating agents. The cell shrinks, then slowly re-expands as the agents enter (you saw this on page 45). This takes several minutes, depending on the cell type and temperature.
- Vitrification solution: a high concentration of permeating agents plus sugar. The cell shrinks quickly. Exposure is short, about a minute in most protocols, and must end with the sample in liquid nitrogen.
Follow the commercial kit's or your laboratory's validated timings exactly. They depend on temperature: a protocol written for room temperature behaves differently on a warmed stage.
Warming solutions
Warming reverses the sequence. The carrier goes straight from liquid nitrogen into a warmed solution with a high sugar concentration, which allows very fast warming (page 45). Then the cell passes through solutions with less and less sugar, so the permeating agents leave and water returns gently. The timing of each step is in the warming protocol (page 52).
Carrier devices
A carrier holds the cell in a tiny drop during cooling and storage. Each has a handle, which is labelled, and a tip, which carries the drop.
- Open carriers: a thin film, strip or loop on which the drop sits and is plunged directly into liquid nitrogen. They give the fastest cooling. After vitrification a protective cap or sleeve is fitted under liquid nitrogen.
- Closed carriers: the cell is loaded into a fine straw or capillary, or onto a strip that is sealed into an outer straw before cooling. The sample never touches liquid nitrogen. Sealing must be complete; a poor seal lets nitrogen in, which can crack the straw on warming.
Each laboratory chooses one validated system for each cell type and trains everyone on it. Never mix carriers and kits that were not validated together.
At the bench
- Prepare equilibration and vitrification solutions in a dish as your protocol states, at the stated temperature, labelled.
- Pre-label carriers with the patient identity and sample details, with a witness, before any cell enters a solution.
- With beads or training material, practise loading the smallest possible drop on the carrier tip under the stereo microscope, then remove excess solution.
- Practise placing caps and sealing closed straws under supervision, then inspect the seal.
Common errors
- Moving a cell into vitrification solution before equilibration is complete.
- Leaving a cell in vitrification solution too long while you look for the carrier.
- Too large a drop on the carrier.
- Labelling carriers after loading instead of before.
- A poor seal on a closed straw.
Key points
- Permeating cryoprotectants enter the cell; non-permeating sugars stay outside and draw water out.
- Cryoprotectants are toxic in high concentration and at warm temperatures, so exposure is short and timed.
- Vitrify in two stages: the equilibration solution first, then the vitrification solution.
- Warming moves the cell through less and less sugar, so water returns gently.
- Label carriers before any cell enters a solution, and never mix carriers and kits that were not validated together.
Check yourself
Name three permeating and two non-permeating cryoprotectants.
Permeating, any three of: ethylene glycol, dimethyl sulfoxide (DMSO), propylene glycol (1,2-propanediol) and, for sperm, glycerol. Non-permeating: sucrose and trehalose.
Why is vitrification done in two stages?
A cell dropped straight into full-strength vitrification solution would shrink too fast, and the cryoprotectant would not have time to enter. So the equilibration solution comes first, then the vitrification solution.
What two jobs does sucrose do in vitrification and warming?
During vitrification it draws water out of the cell by osmosis, so less water is left to freeze. On warming it slows the rush of water back into the cell while the permeating agent leaves, so the cell does not swell until it bursts.
What is the main difference between an open and a closed carrier?
On an open carrier the drop is plunged directly into liquid nitrogen. In a closed carrier the cell is sealed in before cooling, and the sample never touches liquid nitrogen.
From page 45: What are "solution effects"?
When ice forms outside the cell, the water that freezes leaves the salts behind. The remaining liquid becomes very concentrated, pulls water out of the cell and can damage proteins and membranes (page 45).
From page 46: Why does vitrification need such a small volume?
A very small volume speeds cooling (page 46).
Questions on this page
Why are two permeating cryoprotectants combined?
Mixing two agents gives the protective effect of a high total concentration while keeping each agent lower, which reduces toxicity.
Why is sucrose used in warming solutions?
It holds water outside the cell while the permeating cryoprotectant leaves, so the cell does not swell too quickly and burst.
When are carriers labelled?
Before any cell enters a solution, with a witness check. Labelling after loading risks a wrong label and delays the sample.
The Genesis ART Textbook is teaching text for trainees and doctors of the Genesis ART Training Academy. It is not treatment advice and not a laboratory standard operating procedure; every technique is learned under supervision. Page numbers never change.