Daily QC and lab KPIs (Vienna consensus)

On this page you will learn

Three ideas, one aim

All three serve one aim: every patient's gametes and embryos get the same reliable care, every day.

The daily checks

Each morning, before any patient work, the laboratory records a set of readings. Typical checks include:

Every reading is written in the log with the time and your initials. A reading out of range is acted on before any work in that equipment, and the action is recorded.

The Vienna consensus

In 2017, ESHRE and Alpha (Scientists in Reproductive Medicine) published the Vienna consensus on ART laboratory performance indicators. It sets out a group of indicators that cover the main steps of laboratory work. Examples include the proportion of oocytes damaged by ICSI, normal fertilisation after ICSI and after conventional IVF, cleavage on Day 2, blastocyst development, survival of embryos after warming, and implantation.

For each indicator it suggests a "competency" value, which every laboratory should reach, and a "benchmark" level, which is an aspirational target.

It also separates three kinds of indicator: reference, performance and outcome indicators. A reference indicator, such as the proportion of mature oocytes at pick-up, depends on the clinic as well as the laboratory. Performance indicators mainly reflect laboratory work. This helps you work out where a problem truly lies.

Using KPIs well

At the bench

1. Measure; write the reading2. Out of range? Repeat once3. Still out: tell the senior4. Move dishes to a back-upif needed5. Record what you did
Figure: what to do with a QC reading.
  1. Arrive in time to complete the QC checks before the first case.
  2. Measure; do not copy yesterday's number. Write the actual reading.
  3. If a reading is out of range, repeat it once with care. If still out, inform the senior embryologist, move dishes to a back-up if needed and record what you did.
  4. Enter outcomes for each case (fertilisation, development, cryo) on the day, so KPI data are complete.
  5. Take part in the monthly KPI review and ask how your own work compares with the laboratory's competency values.

Quality systems are also expected under the ART (Regulation) Act 2021 and the ICMR national ART guidelines. The records you keep are therefore part of the laboratory's legal duty, not only good habit.

Common errors

Key points

Check yourself

  1. What is the difference between quality control and quality assurance?

    Quality control is checking that equipment and materials are within range today. Quality assurance is the whole system that makes good work likely: protocols, training, records, audits and corrective actions.

  2. Name five readings taken before patient work begins each morning.

    Any five of:

    • incubator temperature
    • incubator carbon dioxide (and oxygen in reduced-oxygen incubators)
    • culture medium pH
    • heated stages, warm blocks and workstation surfaces
    • gas cylinder pressures
    • refrigerator temperatures
    • room temperature and humidity
    • liquid nitrogen levels and the oxygen alarm.
  3. What is the difference between a competency value and a benchmark value in the Vienna consensus?

    A competency value is a level every laboratory should reach. A benchmark is an aspirational target.

  4. Why is the proportion of mature oocytes a reference indicator and not a pure laboratory performance indicator?

    Because it depends on the clinic as well as the laboratory. A performance indicator mainly reflects laboratory work.

  5. From page 17: How is the carbon dioxide set point of an incubator chosen?

    It is chosen to give the pH that the medium maker specifies (page 17).

  6. From page 19: Why are culture dishes usually prepared the afternoon before use?

    So that pH and temperature are stable by morning (page 19).

Contents and reading plan

Questions on this page

Are KPIs figures to show patients?

No. KPIs measure how well each step of laboratory work is done, so problems can be found and fixed. They are internal quality tools.

What if a QC reading is out of range?

Repeat it carefully, inform the senior embryologist, protect the dishes and record the reading, the cause and the action taken.